UV Absorbance and Peptide Concentration: A280 and A205
Peptide AnalysisUV AbsorbanceLab Handling

UV Absorbance and Peptide Concentration: A280 and A205

V8 Peptides Research TeamUpdated

Compiled from peer-reviewed literature and manufacturer analytical data for laboratory research reference.

A peptide's concentration in solution can be read from its UV absorbance with the Beer-Lambert law, A = ε x c x l: at 280 nm for sequences that contain tryptophan or tyrosine, using ε280 = 5500 x (Trp) + 1490 x (Tyr) + 125 x (cystine) M-1cm-1, and near 205 nm, where the peptide backbone absorbs, for sequences that do not.[2] Both readings count peptide actually dissolved, which is not the same number as the weighed mass divided by the volume. This page sets out the formulas, three worked calculations (arithmetic only, with illustrative numbers), what the published validation of the 205 nm method found, and where each method fails. It is a laboratory reference for research use only.

The 280 nm method

Absorbance at 280 nm comes from tryptophan and tyrosine side chains and, to a small extent, from disulfide bonds.[1] Pace and colleagues fitted 116 measured values for 80 proteins and gave the folded-protein value in water as [2]

ε280 (M-1cm-1) = (#Trp) x 5500 + (#Tyr) x 1490 + (#cystine) x 125

and noted that these values are quite reliable for proteins that contain Trp and less reliable for proteins that do not, and that measuring the coefficient is better than predicting it when that is possible.[2] The method they describe as best for measuring it goes back to Edelhoch and was set out for sequence-based calculation by Gill and von Hippel, who report accuracy of +/- 5% in most cases for normal globular proteins.[2][3] Two cautions follow for short peptides. The fit was made on folded proteins, and Pace found the average Trp and Tyr coefficients inside proteins differ from those measured in water, propanol, 6 M guanidine hydrochloride or 8 M urea.[2] None of the sources opened for this page tested the formula on a panel of short synthetic peptides, so its error for a given peptide is not established here.

Worked example 1: 280 nm

Take a sequence with one Trp, one Tyr, no cystine and a molar mass of 1500 g/mol. ε280 = 5500 + 1490 = 6990 M-1cm-1. A 1 cm cuvette reads A280 = 0.699 after blanking on the solvent.

  • Molar concentration: c = A / (ε x l) = 0.699 / (6990 x 1) = 1.00 x 10-4 M, which is 100 µM.
  • Mass concentration: 100 µM x 1500 g/mol = 0.15 g/L = 0.15 mg/mL.
  • If the sample was diluted before reading, multiply by the dilution factor.

A sequence with no Trp, Tyr or cystine returns a result of zero here, which is a limit of the method, not a measurement of an empty vial.

The 205 nm method

Absorbance at 205 nm arises mainly from the peptide backbone, so it can be used for proteins and peptides that lack Trp and Tyr, and it is far more sensitive: the ratio A205/A280 is about 30 on average and varies widely.[1] A generic value of ε205 = 31 mL mg-1cm-1 is often used. Side chains still matter, because the aromatic residues and histidine absorb more than a single peptide bond at 205 nm, and Scopes proposed correcting the value with the A280/A205 ratio: ε205 = 27.0 + 120 x (A280/A205) mL mg-1cm-1.[1]

Worked example 2: 205 nm with the generic value

A reading of A205 = 0.62 in a 1 cm cell, after subtracting the blank, gives c = 0.62 / 31 = 0.020 mg/mL. The same number held in the molar form is ε205 (M-1cm-1) = 31 x molar mass, which is 46,500 for the 1500 g/mol peptide above.

What the validation measured

Anthis and Clore measured six proteins and two short peptides at 205 and 280 nm, taking the 280 nm value as the reference, and compared three ways to get ε205. They proposed a method that sums contributions from the backbone and from each side chain, directly from the sequence, and reported that it needs no calibration.[1] Their results for the two peptides (a one-Trp, 22-residue fragment and a 26-residue one-Trp peptide, both with predicted ε280 of 5500) are in the table; the generic value was off by about a sixth to a fifth.[1]

Peptide (mass, g/mol)ε205 from A280 (mL mg-1cm-1)Generic 31 (error)Scopes method (error)Sequence method (error)
CaMK1 299-320 (2599)37.2131 (-16.7%)33.82 (-9.1%)37.82 (+1.7%)
skMLCK peptide (2964)38.2531 (-19.0%)32.82 (-14.2%)37.76 (-1.3%)

Across all the samples in the study, the generic value was off by up to about 19%, and the sequence-based method by up to about 8%.[1] The generic value works best where the sequence has few aromatic residues: for calmodulin, which has two Tyr and no Trp in 148 residues, it was within about 1% to 5%.[1] A short peptide with a Trp is exactly the case where it undershoots.

Why buffers matter at 205 nm

Most common buffers are essentially transparent at 280 nm, but many solutes absorb at 205 nm, so any buffer absorbance must be measured and subtracted.[1] In that study the undiluted stock buffers read A205 from 0.59 to 4.37 for most proteins and about 23.7, 27.5 and 31.2 for three others, which is why the authors diluted samples 1:1000 to 1:10,000 in water and subtracted a matched buffer blank.[1] The practical rule from the paper: read a sample whose absorbance is much larger than the blank's, and at the least make the difference significantly larger than the experimental error.[1] A peptide dissolved in water needs no correction for the buffer, but still needs a matched blank of the same water and cuvette.

Worked example 3: comparing the reading with the weighed mass

Suppose 5 mg of a lyophilized peptide (molar mass 1500 g/mol, ε280 = 6990) is dissolved in 2.5 mL. The weighed-mass concentration is 5 / 2.5 = 2.0 mg/mL. A 1:10 dilution gives 0.200 mg/mL, which is 0.200 / 1500 = 133.3 µM, so the A280 expected if everything weighed were peptide is 133.3 x 10-6 x 6990 = 0.932. If the cuvette reads 0.746, the ratio is 0.746 / 0.932 = 0.80, so the solution holds 80% of the peptide that the weighed mass implies, or 1.6 mg/mL before dilution.

This gap is the same quantity as net peptide content. The remaining 20% of a lyophilized powder's mass can be water, counter-ion and other non-peptide material, as set out in the net peptide content article, and the reading by itself does not say which. It also assumes the extinction coefficient is right, so a 5% error in ε moves the apparent content by 5%.[3] UV is one of several ways to measure peptide content; a comparison of HPLC assay, qNMR and amino acid analysis on oxytocin found the HPLC assay against a same-material standard gave the lowest inter-laboratory variability.[4] Reviews list quantitative amino acid analysis, nitrogen analysis, UV and qNMR among the absolute content methods.[5]

Where UV concentration goes wrong

  • No chromophore. A sequence with no Trp, Tyr or cystine has no 280 nm signal; use 205 nm and its blank discipline.[1]
  • Wrong coefficient. The 280 nm prediction is described as less reliable without Trp and was fitted on folded proteins.[2] A measured coefficient, from a standard of known content, beats a predicted one.
  • Blank mismatch at 205 nm. Buffer absorbance reached 31 at 205 nm in the validation study, large enough to bury a dilute sample.[1]
  • Counting the wrong thing. UV counts every absorbing species that is present, including degraded or modified forms that still absorb at the same wavelength. Chemical routes are in the aggregation and degradation routes article, and the HPLC purity article covers how separation addresses that.
  • Light scattering. Particles or aggregates scatter light and add apparent absorbance in a UV reading, so a cloudy solution should be assessed before it is quantified (cloudy vial after reconstitution). This is general optics, not a result from the sources above.

Using the numbers with the calculators

The weighed-mass side of the arithmetic, label mass divided by diluent volume, is covered in the volume calculations article and the calculator. The UV side is an independent check of the same solution. Record the wavelength, path length, blank, dilution factor and the source of the coefficient with the result, so the figure can be recomputed later.

Frequently asked questions

Can a peptide without tryptophan or tyrosine be measured at 280 nm?

No useful signal is expected. The 280 nm absorbance of a polypeptide comes from tryptophan and tyrosine residues and, to a small extent, from disulfide bonds, so a sequence with none of them (and no cystine) gives a reading near zero. The peptide-bond absorbance near 205 nm is the usual alternative.

Why use 205 nm if 280 nm is simpler?

Absorbance at 205 nm comes mainly from the peptide backbone and is about 30 times larger than at 280 nm on average, so it works for sequences without aromatic residues and for dilute samples. The cost is that many buffer components also absorb there, so a matched blank is needed.

Does a UV reading confirm what is on the label?

No. It reports how much absorbing peptide is in the solution being read, assuming the extinction coefficient is right. Comparing that figure with the weighed mass divided by the volume is how a mass-versus-content gap shows up; the reading alone does not say which part of the gap is water, counter-ion or something else.

Does V8 publish UV concentration data?

This page explains the method for laboratory planning and makes no statement about V8 reports. See the lab reports page for what V8 publishes.

Related reading

Research use only. Not for human or veterinary use.

References

  1. Anthis NJ, Clore GM. Sequence-specific determination of protein and peptide concentrations by absorbance at 205 nm. Protein Sci 2013;22:851-858. PubMed 23526461; doi:10.1002/pro.2253
  2. Pace CN, Vajdos F, Fee L, Grimsley G, Gray T. How to measure and predict the molar absorption coefficient of a protein. Protein Sci 1995;4:2411-2423. PubMed 8563639; doi:10.1002/pro.5560041120
  3. Gill SC, von Hippel PH. Calculation of protein extinction coefficients from amino acid sequence data. Anal Biochem 1989;182:319-326. PubMed 2610349; doi:10.1016/0003-2697(89)90602-7
  4. Li C, Bhavaraju S, Thibeault MP, et al. Survey of peptide quantification methods and comparison of their reproducibility: a case study using oxytocin. J Pharm Biomed Anal 2019;166:105-112. PubMed 30640042; doi:10.1016/j.jpba.2018.12.028
  5. Yang EJ, Kim SH, Kim A, Choi J, Jeong HJ, Na DH. Regulatory and analytical considerations for the quality assessment of peptide drugs. J Pharm Investig 2026;56:1239-1261. doi:10.1007/s40005-026-00817-2