Dilute peptide solutions lose peptide to the walls of the vials, tubes and pipette tips that hold them, and the loss is large enough to change a measured concentration. In one study, 1 µM solutions of three cationic peptides kept for one hour in borosilicate glass vials or polypropylene tubes returned only 10% to 20% of the expected amount, and four successive transfers between such containers left close to nothing.[1] Which surface loses least depends on the peptide: in another study the same plastic gave 90% recovery for one peptide and 11% for another.[2] This page sets out what each published test measured, a worked surface-capacity calculation, what reduced the loss, and where the evidence stops. It is a laboratory reference for research use only.

What the published tests measured

StudyPeptides and set-upMain finding
Kristensen 2015[1]Mastoparan X, melittin, magainin 2; 220 µL or 2 mL in borosilicate glass, polypropylene, Protein LoBind tubes and quartz; analysis by HPLCRecovery of 10% to 20% at 1 µM in glass and polypropylene, rising with concentration; LoBind lost much less; adsorption took seconds
Goebel-Stengel 2011[2]Eight radiolabelled peptides (ghrelin, CCK-8S, insulin, PYY, leptin, GLP-1, CRF, nesfatin-1), 2 mL, 48 h at 4 C, six tube typesBest surface differed by peptide; recovery without BSA ranged from 3.7% to 90.3%
Karvanen 2017[3]Colistin 0.125 to 8 mg/L in glass, polypropylene, polystyrene, low-protein-binding microtubes, microplates, over 24 hStart concentrations 44% to 102% of expected in large tubes; loss over 24 h; low-protein-binding polypropylene lost least
Grohganz 2004[4]Cetrorelix at 0.2 to 4 µg/mL in vials of different materials and media (abstract read)Langmuir plateau 0.4 µg/cm2; glass above polypropylene and polyethylene above PTFE; differences between glass vials of different suppliers
Judák 2017[5]GHRP-5, TB-500, insulin lispro, Synachten on various surfaces (abstract read)Low-bind consumables were not beneficial in all cases; choose by the peptide's properties
Stejskal 2013[6]Tryptic digest, 1 fmol per protein, six sample solutions, seven vial surfaces (abstract read)Vial material mattered less than the sample solution; a polypropylene vial with 0.001% PEG, or 6 M urea with 1 M thiourea, both with 0.1% formic acid, did best

Concentration: the same tube loses a larger share at low concentration

A container wall can bind only so much peptide. When the total amount in the tube is small compared with that capacity, most of it ends up on the wall; when it is large, the same wall takes a smaller share. In Kristensen's tubes the recovered percentage rose with peptide concentration, which the authors read as the walls saturating.[1] Cetrorelix on vial surfaces followed a Langmuir isotherm that reached a plateau at 0.4 µg/cm2, and colistin in large tubes was fitted by the same model.[3][4]

The arithmetic below is illustrative, not a measurement. Assume a tube of 9 mm inner diameter holding 220 µL, as in the Kristensen tubes. The liquid is 3.46 mm deep and touches 1.61 cm2 of wall and floor. At the cetrorelix plateau of 0.4 µg/cm2 that surface could take up to 0.65 µg. Mastoparan X (INWKGIAAMAKKLL-NH2) has a calculated average mass of 1556 Da, so the tube holds this much peptide:

Concentration (µM)Peptide in 220 µL (µg)Share the wall could take if filled to 0.4 µg/cm2
10.340%
20.686%
51.7162%
103.4281%
206.8591%

The shape matches the paper: nearly everything is at risk at 1 to 2 µM and the share falls as concentration rises. The numbers do not match exactly, and should not be expected to. The measured recovery at 1 µM was 10% to 20%, not zero, because the capacity came from a different peptide, the geometry is idealised, and real capacity depends on the surface.[1][4]

Surface-to-volume ratio, transfers and dilution steps

  • Larger volume, smaller share lost. For 2 µM solutions, a 2 mL volume gave a higher recovery than 220 µL in glass and polypropylene, although significant loss remained in some cases. In quartz cuvettes with a stirring bar, recovery was 50% or less for all three peptides.[1]
  • Each transfer is a new surface. Moving 250 µL of a 5 µM solution through four borosilicate glass or polypropylene containers, with an hour in each, left close to 0% of the peptide. Protein LoBind tubes lost much less.[1]
  • Dilution series lose peptide at every step. For colistin, the loss per dilution step ranged from 0% to 57% with an incremental scheme and from 0% to 80% with a serial one, and the authors name three drivers: the number of dilution steps, the materials, and the concentration.[3]
  • Time matters, but not always. For mastoparan X, adsorption to glass and polypropylene was complete within seconds, and 10 s, 1 h and 24 h gave similar recoveries.[1] For colistin the loss continued over hours, with half-lives of 0.9 to 11.6 h depending on material and concentration.[3]

Material choice: no single best surface

In the radiolabelled-peptide study, each peptide was incubated at about 5 fmol in 2 mL for 48 hours at 4 C in six tube types. Best and worst recoveries for each:[2]

PeptideBest surface (recovery)Worst surface (recovery)
GhrelinPolypropylene (90.3%)Flint glass (20.0%)
CCK-8SPolypropylene (86.3%)Polystyrene (16.0%)
InsulinPolyallomer (68.3%)Polypropylene (10.9%)
PYYBorosilicate glass (67.7%)Polystyrene (9.3%)
LeptinBorosilicate glass (60.3%)Polystyrene (21.9%)
GLP-1Borosilicate glass (39.6%)Polystyrene (3.7%)
CRFBorosilicate glass (61.5%)Polystyrene (55.8%)
Nesfatin-1Borosilicate glass (32.0%)Polycarbonate (27.4%)

The same paper found borosilicate glass equal to or better than flint glass, and polypropylene equal to or better than polystyrene, for these peptides. Net charge, size and hydrophobicity did not predict the pattern: the authors state that no rule could be drawn from their data and that the results should not be applied to other peptides.[2] Siliconised tubes, often used to cut peptide loss, lowered recovery for most of the peptides (by 53% for PYY, 49% for insulin and 46% for CCK-8S) and helped only nesfatin-1 and CRF in polycarbonate.[2]

Other studies fit the same message. Colistin was lost least in low-protein-binding polypropylene, then polypropylene and glass, and most in polystyrene, so the authors advise against polystyrene.[3] For the four peptides in the doping-control study, expensive low-bind consumables were not beneficial in every case, and the authors recommend choosing by the peptide's physical and chemical features.[5] In Kristensen's data, magainin 2 kept a lower share in a 2 mL LoBind tube than in a 220 µL one, even though the other two peptides did not.[1]

What reduced the loss in published tests

MeasureWhat was foundCaveat from the same source
Higher concentration, larger volume, fewer transfersLess relative loss as the wall saturates and the ratio of surface to volume falls[1][3]Cannot be used if the experiment needs a low concentration
Carrier protein, 1% BSARecovery of the eight radiolabelled peptides rose to 85% to 98% on their best surface; freeze-drying with BSA in the best tube kept more than 89%[2]Not feasible where the sample goes on to LC-MS analysis[2]
Lipid vesicles1 mM POPC/POPG vesicles kept mastoparan X, melittin and magainin 2 off the walls[1]Only for experiments that already contain membranes
Solvent and surfactant choiceFor cetrorelix, ethanol, an acidic pH from acetic acid, or a micellar surfactant raised the HPLC response up to fivefold[4]Peptide-specific; abstract read
Sample solution for autosamplers0.001% PEG or 6 M urea with 1 M thiourea, with 0.1% formic acid, preserved more peptides than other solutions[6]Digest peptides at femtomole level; abstract read
Antiadsorption diluentA diluent made by controlled hydrolysis and precipitation of BSA lowered adsorption of some peptides to glass and works with LC-MS[7]"Certain peptides"; abstract read
Presaturating the containerReported to help in earlier work[1]For mastoparan X, adsorption was partly reversible, so a presaturated wall can release peptide later[1]
Salt concentrationNo significant effect for the three cationic peptides at 2 to 149 mM NaCl, apart from a slight trend in glass[1]An earlier report found an effect for salmon calcitonin[1]

A 2014 review of strategies in LC-MS bioanalysis concludes that, because peptides differ so much in physicochemical properties, no general approach to minimising adsorption exists.[8]

A recovery check that a lab can run

Kristensen's design is a reusable template: prepare the working solution in the container under test, hold it for the time the experiment needs, transfer a fixed volume, and compare the measured amount with a standard prepared with minimal surface contact, for example one made from a concentrated stock directly in the analysis vial with an additive that keeps the peptide off the wall (Kristensen used lipid vesicles). The ratio is the recovery. The standards in that paper were made from a 100 µM stock handled in low-bind tubes, so the pipette tips were already saturated, and the authors ran a control to rule out tip loss before attributing the loss to the tubes.[1] Repeat at the lowest concentration used in the real work, because that is where the loss is largest. Concentrations computed from a weighed stock, as in the reconstitution calculator, are nominal; they describe what was added, not what stayed in solution.

Limits of the evidence

  • Few peptides, specific systems. Three membrane-active cationic peptides, eight iodinated hormones at femtomole levels, a cationic lipopeptide antibiotic and a tryptic digest are not a sample of all peptides. None of these studies tested the peptides V8 sells, and none of them tested a product vial.
  • Iodinated tracers. The authors of the radiolabel study note that iodination could change binding, though the wide spread between peptides argues against it being the cause.[2]
  • Access. Kristensen, Goebel-Stengel and Karvanen were read in full text. The Grohganz, Judák, Stejskal, Verbeke and Maes papers were read as abstracts only, so their methods are not reported here.
  • Manufacturers differ. Glass from different suppliers behaved differently for cetrorelix, and a product name such as "low-bind" covers different surface treatments.[3][4]

Frequently asked questions

Which tube or vial is best for peptide solutions?

No material was best across the published tests. In one set of eight peptides, polypropylene gave the highest recovery for two, borosilicate glass for five and polyallomer for one, and the worst surface for each peptide cost between about 5 and 70 percentage points. The authors could find no rule that predicts the best surface from charge, size or hydrophobicity.

Do low-bind tubes solve peptide loss?

They reduced it for three cationic peptides in one study and for colistin in another, but not in every case: a doping-control study of four peptides found that low-bind consumables were not beneficial in all cases, and one of the three cationic peptides still lost more in a 2 mL low-bind tube than in a 220 µL one.

Does diluting a peptide in several steps lose more?

In the colistin study, each step that put the solution on fresh plastic lost some, from 0 to 57% per step in an incremental scheme and 0 to 80% in a serial one, and the loss grew as the concentration fell. The authors advise minimising dilution steps when the medium has no protein.

Related reading

For the preparation side see how to reconstitute a lyophilized peptide, reconstitution volume calculations and aliquot planning; for what else changes a stored solution, pH and peptide stability and solubility and solvent choice.

Research use only. Not for human or veterinary use.

References

  1. Kristensen K, Henriksen JR, Andresen TL. Adsorption of cationic peptides to solid surfaces of glass and plastic. PLoS One 2015;10:e0122419. PubMed 25932639; doi:10.1371/journal.pone.0122419
  2. Goebel-Stengel M, Stengel A, Tache Y, Reeve JR Jr. The importance of using the optimal plasticware and glassware in studies involving peptides. Anal Biochem 2011;414:38-46. PubMed 21315060; doi:10.1016/j.ab.2011.02.009
  3. Karvanen M, Malmberg C, Lagerbäck P, Friberg LE, Cars O. Colistin is extensively lost during standard in vitro experimental conditions. Antimicrob Agents Chemother 2017;61:e00857-17. PubMed 28893773; doi:10.1128/AAC.00857-17
  4. Grohganz H, Rischer M, Brandl M. Adsorption of the decapeptide Cetrorelix depends both on the composition of dissolution medium and the type of solid surface. Eur J Pharm Sci 2004;21:191-196. PubMed 14757490; doi:10.1016/j.ejps.2003.10.008
  5. Judák P, Van Eenoo P, Deventer K. Adsorption effects of the doping relevant peptides Insulin Lispro, Synachten, TB-500 and GHRP 5. Anal Biochem 2017;537:69-71. PubMed 28887173; doi:10.1016/j.ab.2017.09.003
  6. Stejskal K, Potěšil D, Zdráhal Z. Suppression of peptide sample losses in autosampler vials. J Proteome Res 2013;12:3057-3062. PubMed 23590590; doi:10.1021/pr400183v
  7. Verbeke F, Bracke N, Debunne N, Wynendaele E, De Spiegeleer B. LC-MS compatible antiadsorption diluent for peptide analysis. Anal Chem 2020;92:1712-1719. PubMed 31874035; doi:10.1021/acs.analchem.9b01840
  8. Maes K, Smolders I, Michotte Y, Van Eeckhaut A. Strategies to reduce aspecific adsorption of peptides and proteins in liquid chromatography-mass spectrometry based bioanalyses: an overview. J Chromatogr A 2014;1358:1-13. PubMed 25022477; doi:10.1016/j.chroma.2014.06.072